M1322多功能数据记录仪 - 温湿度、CO2浓度计,2路MiniDIN输入与2路通用输入通道
通用型数据记录仪,4路输入通道,内置CO2传感器
多功能数据记录仪可固定长期连接以太网使用或作为便携式设备使用
产品附有一份可追溯校准证书,基于ISO/IEC17025标准。
记录仪可用于测量温度、湿度、CO2浓度(内置传感器),记录时间1秒-24小时可调。
数据记录仪具备以太网接口,接入以太网可实现网页在线监测数据,并可将数据导出并保存。
- 在线监测
- 数据记录
- 报警指示
-
型号M1322
-
保3 年
技术参数
测量参数
产品特点
Applications:
Healtcare and laboratories - Monitoring of storage areas at temperatures up to -200 °C.
Food industry - Monitoring of critical processes according to the HACCP standard with the possibility of prompt notification of unforeseen events that could lead to the degradation of the goods.
Industry and production - monitoring temperature and humidity in food, pharmaceutical, aerospace etc.
Production facilities and warehouses - Monitoring of storage conditions and production processes in the temperature range -200 ° C to + 600 ° C. Sending data to COMET Database or SCADA third-party systems.
Metrology - COMET sensors together with COMET radiationshields are the ideal basis for weather stations for long-term weather, climate and air monitoring.
Data centers, server rooms - Temperature and humidity monitoring, data center monitoring, RACK stands, including flood status reports, opening / closing of doors (windows), motion, smoke. SNMP monitoring, remote alarms by email and Syslog.
Building management - Our instruments which measure temperature, humidity, atmospheric pressure and CO2 are an integral part of these industry-leading solutions.
Features description:
● the features may vary depending on the device type● memory for 1 million values
● indication of alarm conditions via e-mail, SMS, integrated acoustic siren LEDs, stwo-state output to control other devices
● analytical software for data processing and configuration - COMET Vision
● communication interfaces - USB, RS232 and Ethernet (SNMPv1 and XML)
Up to 4 physical Inputs - 2x TERMINALS + 2x MiniDin
- Sensors with woltage output (range 0-10 V; accuracy ±10 mV)
- Sensors with current output (range 0-20 mA; accuracy ±20 µA)
- Devices with two-state output (potential-less contact or voltage signal 0-30V); may be applied to input 3 and 4
- Devices with pulse signal (potential-less contact or voltage signal 0-30V); may be applied to input 4
- Temperature probes Pt1000 with built-in configuration memory (serie xxx/M)
- Temperature/relative humidity probes with digital output - DigiS/M, DigiL/M, DIGIH/M
- The probes are interchangeable without calibration to a specific device and regardless of the length of the cable. The length of the cable can be 1, 2, 5, 10, 15 meters
- Internal CO2 sensor with range 0-2000 ppm (custom range 0-10.000ppm)
Evaluation of up to 16 variables
The measured values are stored in the internal non-volatile memory. Memory capacity is up to 1 million values. Logging mode can be adjusted as non-cyclic, when logging stops after filling the memory, or cyclic, when after fulfilling the memory oldest stored values are overwritten by new ones.
Measured values are displayed on the screen with backlight option and the ability to change the size of display digits
Power supply
Multilogger can be powered by AC adapter or by rechargeable batteries with charging capability. Batteries are always a part of the device (you can also use standard alkaline AA batteries). Batteries power the device for several months.
While using the AC adapter, the batteries provide backup operations and provide recording and measurement during power failure. Thanks to the Ethernet output the device informs about the failure via e-mail message.
Alarm limits and alarm indication
It is possible to set two independent alarm limits for each channel (i.e. measured or calculated value).
Alarm signalization can be acoustic (built-in beeper), optical, sending an e-mail alert or an alarm output. The alarm output is for connecting an external audio or visual alarm.
Communication interfaces
Multilogger is always equipped with USB connector for communication, data analysis and also for charging of the rechargeable batteries.
Included Ethernet interface allows to:
- send an e-mail when an alarm state starts or ends
- use DATALINK: display current values or download values from memory to COMET Database in your PC. Software COMET Database contains many useful tools for data analysis - graphs, tables, statistics, etc.
- simple wireless communication with Wi-Fi TP-LINK router
- third-party applications to read the actual measured values using universal protocols SNMPv1 and XML
- view current measured values using your web browser
Output ALARM OUT
- Output can be used for controling other devices when alarm is launched, such as external buzzer, telephone dialer etc.
- Type of output - open collector transistor
- Max. switching current - 100 mA
- Max. voltage on output - 24 V
- Auxiliary voltage at terminal - +5 V (only when the mains supply is present)
- Users can receive alarms if the values go outside the paramters set. Alarms are indicated through an internal buzzer , 3 LEDs or email alerts via Ethernet port.